CRDS Trace Moisture Analyzer with Innovative Spectral Analysis Technology Achieving Unprecedented Miniaturization

CRDS Trace Moisture Analyzer
DewTracer mini CRDS-H₂O

DewTracer mini CRDS-H₂O is a trace moisture analyzer for a wide range of gases, based on CRDS (Cavity Ring-Down Spectroscopy) technology utilizing laser techniques.
With proprietary optical design and spectral analysis technology, it achieves unprecedented miniaturization.
It enables highly sensitive, highly accurate, and high-speed measurement of trace moisture (12 ppbv to 20 ppmv, -100 to -55°C DP) in high-purity gases, including carrier gases and atmospheric gases.

* DewTracer mini CRDS-H2O has been developed with cooperation of the National Institute of Advanced Industrial and Technology (AIST).

Brochure(English

High-Speed Response

1000 ppbv → 50 ppbv
63% Response 15.4 s
90% Response 52.4 s
(Measured Gas : N₂)

50 ppbv → 1000 ppbv
63% Response 26.5 s
90% Response 37.4s
(Measured Gas : N₂)

High-Accuracy

Comparison Data with Primary Standards provided by the National Institute of Advanced Industrial Science and Technology (AIST) in Japan (1)(Measured Gas : N₂)

Comparison Data with Primary Standards provided by the National Institute of Advanced Industrial Science and Technology (AIST) in Japan (2)(Measured Gas : N₂)

High-Sensitivity

Comparison Data with Primary Standards provided by the National Institute of Advanced Industrial Science and Technology (AIST) in Japan (3)(Measured Gas : N₂)

Reliable Measurement Data

  • Uncertainty is calculated through collaborative research with the National Institute of Advanced Industrial Science and Technology (AIST) in Japan.
  • All units shipped undergo comparative calibration against internal reference instruments traceable to national standards supplied by AIST.

User-Friendly Design

Selectable Installation Options Based on Applications and Locations

Simple Operation

Various Functions Available via Optional Monitor (Equipped with a 5-inch Full-Color Touch Panel) or a PC with Bundled Software Installed.

  • Start and stop measurements
  • Configure various settings
  • Display settings (Measurement Values, Trend Graphs, Spectral Waveforms)
  • Malfunction prevention lock function
  • Sensor operation status monitoring
  • Alarm function and more…

External Data Output

Measurement values or calculated values can be output externally in various formats.
* An optional monitor is required.
* Digital output (RS-232C) comes standard with the optional monitor.
Analog output and relay output are available as additional options.

Safety Measures

  • UJR/VCR fittings are used for gas-contacting parts, ensuring compatibility with chemical-resistant and corrosive gases.
  • To prevent hazardous gases, such as flammable or toxic gases, from leaking outside the enclosure in the unlikely event of a pipe leak within the housing, the device is equipped with a purge port for inert gases like nitrogen to maintain internal enclosure safety.

Specifications

Product NameCRDS Trace Moisture Analyzer
Product TypeDewTracer mini CRDS-H2O
Measurement MethodCRDS (Cavity Ring Down Spectroscopy)
Measurement Range12 ppbv to 20 ppmv (-100 to -55 ℃ DP)
Accuracy±4% or ±12 ppbv
Response SpeedApprox. 60 s (90% response for 50 ⇔ 1000 ppbv change)
Operating ConditionsTemp.: 10 to 40 ℃
Humidity: 35 to 80% RH (non-condensing)
Sample GasType: Air, N₂, O₂, Ar, etc.
* Other gases are supported. Contact us for more information and assistance.
Temp.: 5 to 60 ℃
Flow Rate: < 2.0 L/min
Pressure: 86 to 106 kPaA (Options for < 900 kPaA available)
Purge GasFlow Rate: < 5.0 L/min (when using the purge function)
Measured ValuesVolume mole fraction [ppmv]
Gas temp. [℃]
Gas pressure [PaA]
Piping1/4 inch (1/4 inch : UJR-RP)
Power SupplyAC adapter (100-240 V, 50/60 Hz, 1.4 A max)
Power ConsumptionApprox. 10 W
DimensionsW 150 x H 165 x D 300 [mm]
WeightApprox. 5 kg (excluding AC adapter)
Product NameOptional Monitor (for DewTracer mini CRDS-H₂O)
Product Typemini CRDS PC-D0V1
DisplayColor LCD (800×480 WVGA), 5-inch touch panel
Measured ValuesVolume mole fraction [ppmv]
Gas temperature in measurement unit [℃]
Gas pressure in measurement unit [PaA]
Calculated ValuesDew Point / Frost Point Temperature [℃DP]
Output    Digital: RS-232C
Analog: 4 to 20 mA (Optional)
Relay: MOS FET Relay (100 V / 300 mA) (Optional)
Power SupplyPowered by DewTracer mini CRDS-H₂O
(Dedicated power cable included)
Power ConsumptionApprox. 30 W
DimensionsW 150 x H 200 x D 75.4 [mm]
WeightApprox. 1.7 kg

Key Applications

●High-purity gas production, inspection, and filling processes for semiconductors
 → Process efficiency improvement
   Cost reduction through decreased consumption of expensive gases
●R&D and manufacturing processes for all-solid-state batteries
●Management and control of glove boxes, dry chambers

○R&D and manufacturing processes for semiconductors
○R&D and manufacturing processes for high-purity medical gases
○R&D and manufacturing processes for pharmaceuticals
○R&D and manufacturing processes for lithium-ion batteries (LiB)
○R&D and manufacturing processes for electric double-layer capacitors (EDLC)
○R&D and manufacturing processes for organic EL devices
○R&D and manufacturing processes for flat panel displays
○Anti-corrosion and condensation prevention for precision electronic components
○Humidity control in powder manufacturing processes
○Inspection, performance verification, and evaluation of high-barrier films
○Manufacturing processes for high-purity silicon and other semiconductor materials, etc.

Technical Documents