Precisely controlled humidified gas with specific humidity or dew point is supplied externally using the divided-flow method
SRG Series – Humidity Supply System
STG Series – Dew Point Supply System
Precisely conditioned humidified gas is supplied externally using the divided-flow method defined in JIS B 7920: Humidity – Test Methods, Section 5.3.4.
SRG Series controls humidified gas based on relative humidity, while STG Series controls it based on dew point.
Customizations and special solutions designed for specific applications and requirements are also available upon request.
Features
- Precise control of humidity and dew point generated using the divided-flow method.
- Support for custom design and production tailored to specific applications and purposes.
- Applicable to various non-air gases (contact us for details).
Key Applications
- Performance evaluation testing for new materials, electronic components, and sensors
- Research and development in biotechnology, pharmaceuticals, and related fields
- Manufacturing processes requiring precise humidity control, etc.
Product Name | Humidity Supply System |
Model | SRG-1M |
Humidity Generation Method | Divided-Flow Method |
Dry Air Generation Method | Continuous Regeneration System |
Flow Meter | Mass Flow Controller |
Humidity Range | 0 ~ 100 %rh |
Accuracy | ± 2 %rh at 25 ℃ |
Repeatbility | ± 1 %rh |
Maximum Generation Flow Rate | 20 L/min |
Product Name | Dew Point Supply System |
Model | STG-H-AW |
Dew Point Generation Method | Divided-Flow Method |
Water Supply Method | Automatic Water Supply |
Dew Point Range | 35 ~ 95 ℃DP |
Accuracy | 35 ~ 80 ℃DP:± 0.3 ℃DP 80 ~ 95 ℃DP:± 0.5 ℃DP |
Repeatbility | 35 ~ 80℃DP:± 0.15 ℃DP 80 ~ 95℃DP:± 0.3 ℃DP |
Maximum Generation Flow Rate | 80 ℃DP:2.2 L/min max 95 ℃DP:1.6 L/min max |
* The above specifications are examples only.
Custom design and production are available based on specific applications and conditions.